Neuroretinal rim in non-glaucomatous large optic nerve heads: a comparison of confocal scanning laser tomography and spectral domain optical coherence tomography.
Summary
In large optic discs, BMO-MRW is thinner compared with normal-sized discs and correlates better than CSLT parameters with the RNFLT.
Abstract
BACKGROUND/AIMS
To compare margin-based rim area measurements from confocal scanning laser tomography (CSLT) with Bruch's membrane opening (BMO)-based measurements from spectral domain optical coherence tomography (SD-OCT) by analysis of a group of non-glaucomatous eyes with large optic discs, so-called macrodiscs (disc size >2.45 mmin CSLT). Objective is to create a reference base for large optic nerve heads in SD-OCT diagnostics.
METHODS
102 eyes received CSLT and SD-OCT measurements and clinical examination on the same day. Visual field testing confirmed absence of glaucomatous defects. Statistical and correlation analysis was performed for rim area by CSLT as well as retinal nerve fibre layer thickness (RNFLT) and BMO minimal rim width (BMO-MRW) by SD-OCT.
RESULTS
Mean disc size in CSLT was 2.98±0.4 mm(range 2.45-4.23), mean rim area of 1.55±0.4 mm. BMO area was 2.51±0.33 mm(range 1.61-3.51), mean global RNFLT was 79.55±17.2 μm, mean global BMO-MRW was 234.84±48.3 μm. Correlation of BMO-MRW to global RNFLT was stronger (r=0.60, p<10) than correlation of CSLT rim area to global RNFLT (r=0.26, p=0.24). BMO-MRW and CSLT rim area correlated with r=0.59 (p<10). BMO-MRW and RNFLT significantly decreased with increasing age (p<0.001). Annual loss of BMO-MRW was 0.8 μm/year (R=0.14, p<0.001), loss of RNFLT was 0.27 μm/year (R=0.17, p=0.001).
CONCLUSIONS
In large optic discs, BMO-MRW is thinner compared with normal-sized discs and correlates better than CSLT parameters with the RNFLT. An age-depended loss of BMO-MRW needs to be taken into account in evaluation of the neuroretinal rim.
Keywords
More by Philip Enders
View full profile →Novel Bruch's Membrane Opening Minimum Rim Area Equalizes Disc Size Dependency and Offers High Diagnostic Power for Glaucoma.
The novel role of lymphatic vessels in the pathogenesis of ocular diseases.
The use of Bruch's membrane opening-based optical coherence tomography of the optic nerve head for glaucoma detection in microdiscs.
Top Research in Optic Nerve & Disc
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.