A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer.
Nikhil S Choudhari, Suresh Ramavath, Gunturu Rekha, Aparna Rao, Harsha L Rao, Sirisha Senthil, Chandra Sekhar Garudadri
Summary
The screening criteria of calibration error within ±1 mm Hg at the 0 mm Hg testing level seem to be clinically useful to detect GAT with an acceptable calibration error as per the APGS guideline.
Abstract
BACKGROUND
To investigate whether a lower-end calibration error check alone can identify a Goldmann applanation tonometer (GAT) with acceptable calibration error as per Asia-Pacific Glaucoma Society (APGS) and World Glaucoma Association (WGA) guidelines.
METHODS
This was a multicentric prospective study in a tertiary eye-care institutional setting. Seventy-six slit-lamp-mounted Haag-Streit GATs (Model AT 900 C/M) were included in the study. All GATs were checked for calibration error at 0, 20, and 60 mm Hg testing levels as recommended by the manufacturer. The ability of 2 screening criteria (calibration error within ±1 or ±2 mm Hg, respectively, at the 0 mm Hg testing level without using the calibration error check weight bar) to identify a tonometer within acceptable calibration limits defined by APGS (calibration error within ±2, ±3, and ±4 mm Hg at 0, 20, and 60 mm Hg testing levels, respectively) and WGA (calibration error within ±1 mm Hg at all testing levels) was determined.
MAIN OUTCOME MEASURE
The negative likelihood ratio of the screening criteria
RESULTS
: Five (6.6%) and 42 (55%) GATs were outside of the APGS and the WGA calibration limits, respectively. Negative likelihood ratios of the first and the second screening criteria at the APGS definition of calibration error tolerance were 0.11 and 0.25, respectively, and at that of the WGA definition of calibration error tolerance were 0.38 and 0.90, respectively.
CONCLUSIONS
The screening criteria of calibration error within ±1 mm Hg at the 0 mm Hg testing level seem to be clinically useful to detect GAT with an acceptable calibration error as per the APGS guideline.
More by Nikhil S Choudhari
View full profile →Outcomes of Descemet-Stripping Endothelial Keratoplasty in 52 Eyes With Iridocorneal Endothelial Syndrome.
Is Ahmed Glaucoma Valve Consistent in Performance?
Fifteen-Year Incidence Rate of Primary Angle Closure Disease in the Andhra Pradesh Eye Disease Study.
Top Research in Diagnosis & Screening
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Dry eye disease and oxidative stress.
Central Corneal Thickness in the Ocular Hypertension Treatment Study (OHTS).
Discussion
Comments and discussion will appear here in a future update.