Intraday Repeatability of Bruch's Membrane Opening-Based Neuroretinal Rim Measurements.
Philip Enders, Andreas Bremen, Friederike Schaub, Manuel M Hermann, Michael Diestelhorst, Thomas Dietlein, Claus Cursiefen, Ludwig M Heindl
Summary
Assessment of BMO-MRW by SD-OCT shows high intraday repeatability, which is comparable to the parameter RNFL thickness.
Abstract
PURPOSE
To assess possible intraday variability in Bruch's membrane opening-based (BMO) assessment of neuroretinal rim by spectral-domain optical coherence tomography (SD-OCT) of the optic nerve head (ONH) as well as to evaluate its independence from intraindividual IOP changes.
METHODS
In this noninterventional, prospective cohort study, 51 consecutively enrolled patients with glaucoma or ocular hypertension underwent standardized SD-OCT of the ONH and IOP assessment at two different time points with a time gap of 5 hours on the same day. Random effects models, intraclass correlation coefficients (ICC) and Bland-Altman plots were used to analyze repeatability of BMO minimum rim width (BMO-MRW) and area (BMO-MRA) and peripapillary retinal nerve fiber layer (RNFL) thickness measurements.
RESULTS
Mean BMO area was 1.86 ± 0.30 mm2. At baseline, mean BMO-MRW was 206.46 ± 0.86 μm, mean BMO-MRA was 0.89 ± 0.34 mm2, and mean RNFL thickness was 71.61 ± 0.20 μm. Within intraindividual measurements, mean difference was 2.95 μm or 1.76% for BMO-MRW, 0.02 mm2 or 2.68% for BMO-MRA, and 1.18 μm or 1.89% for RNFL thickness. Comparing time points, η2 was 0.001 (P = 0.757) for RNFL thickness, 0.043 (P = 0.01) for BMO-MRA, and 0.07 (P = 0.06) for BMO-MRW. Intraclass correlation coefficients were greater than 0.994, respectively. Variability of morphometric parameters did not correlate to intraindividual IOP fluctuations.
CONCLUSIONS
Assessment of BMO-MRW by SD-OCT shows high intraday repeatability, which is comparable to the parameter RNFL thickness. Variability seems not to be driven by typical moderate intraday fluctuations of IOP. The two-dimensional parameter BMO-MRA incorporates a fairly higher intraindividual variability.
More by Philip Enders
View full profile →Novel Bruch's Membrane Opening Minimum Rim Area Equalizes Disc Size Dependency and Offers High Diagnostic Power for Glaucoma.
The novel role of lymphatic vessels in the pathogenesis of ocular diseases.
The use of Bruch's membrane opening-based optical coherence tomography of the optic nerve head for glaucoma detection in microdiscs.
Top Research in Optic Nerve & Disc
Browse all →Efficacy of a Deep Learning System for Detecting Glaucomatous Optic Neuropathy Based on Color Fundus Photographs.
Relationship between Optical Coherence Tomography Angiography Vessel Density and Severity of Visual Field Loss in Glaucoma.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Discussion
Comments and discussion will appear here in a future update.