Assessment of Circumferential Angle Closure with Swept-Source Optical Coherence Tomography: a Community Based Study.
Natalia Porporato, Mani Baskaran, Tin A Tun, Rehena Sultana, Marcus C L Tan, Joanne H M Quah, John Allen, David S Friedman, Ching-Yu Cheng, Tin Aung
Summary
In this large community-based study, SS-OCT exhibited moderate performance for angle closure detection compared to gonioscopy as the reference standard.
Abstract
PURPOSE
To evaluate the diagnostic performance of swept-source optical coherence tomography (SS-OCT, CASIA SS-1000; Tomey Corporation, Nagoya, Japan) for angle closure detection, in comparison with gonioscopy, in a community setting.
DESIGN
Reliability analysis.
METHODS
A total of 2027 phakic subjects aged ≥50 years, with no previous history of glaucoma, laser (including peripheral iridotomy), intraocular surgery, or ocular trauma, were consecutively recruited from a community polyclinic in Singapore. Gonioscopy was performed by a single trained ophthalmologist. SS-OCT angle scans, which obtain radial scans for the entire circumference of the angle, were analyzed by a single examiner, masked to the subject's clinical details. On SS-OCT images, angle closure was defined as contact between the iris and any part of the angle wall anterior to the scleral spur. Different cutoff values of the degree of circumferential angle closure (≥35%, ≥50%, and ≥75%) were taken for analysis to assess SS-OCT performance in detecting angle closure.
RESULTS
A total of 1857 subjects (91.6%) were included in the final analysis after excluding poor-quality SS-OCT scans. Almost 90% of the subjects were Chinese, with a mean age of 61.8 ± 6.7 years, and more than half were women (63.5%). The overall AUC of SS-OCT manual grading against gonioscopy was 0.84 (95% confidence interval, 0.81-0.88). The prevalence of angle closure on SS-OCT was 26.1% for the ≥35% definition, with an area under the curve of 0.80 (0.77-0.84), sensitivity of 82.5% (75.3%-88.4%), and specificity of 78.5% (76.5%-80.4%). The first-order agreement coefficient statistics for the 2-quadrant gonioscopic definition of angle-closure with corresponding ≥35%, ≥50%, and ≥75% angle closure definitions for SS-OCT were good at 0.89 (0.83-0.93), 0.88 (0.842-0.93), and 0.88 (0.831-0.99), respectively.
CONCLUSIONS
In this large community-based study, SS-OCT exhibited moderate performance for angle closure detection compared to gonioscopy as the reference standard.
More by Natalia Porporato
View full profile →Ocular Biometric Risk Factors for Progression of Primary Angle Closure Disease: The Zhongshan Angle Closure Prevention Trial.
Role of anterior segment optical coherence tomography in angle-closure disease: a review.
Biometric Risk Factors for Angle Closure Progression After Laser Peripheral Iridotomy.
Top Research in OCT & Imaging
Browse all →Optical coherence tomography angiography: A comprehensive review of current methods and clinical applications.
Deep learning in ophthalmology: The technical and clinical considerations.
Anterior segment optical coherence tomography.
Discussion
Comments and discussion will appear here in a future update.