The Relationship Between Corvis ST Tonometry Parameters and Ocular Response Analyzer Corneal Hysteresis.
Takashi Fujishiro, Masato Matsuura, Yuri Fujino, Hiroshi Murata, Kana Tokumo, Shunsuke Nakakura, Yoshiaki Kiuchi, Ryo Asaoka
Summary
CST parameters are significant, but weakly or moderately, related to ORA measured CH.
Abstract
PRECIS
Corvis ST Tonometry and Ocular Response Analyzer (ORA) measurements were conducted in primary open-angle glaucoma and normative subjects. Many parameters were significantly correlated, however, the strengths were weak to moderate.
PURPOSE
Reichert ORA parameters are derived from pressure information following the application of air-jet, whereas detailed structural observation can be made using the Corneal Visualization Scheimpflug Technology instrument (CST). The purpose of the study was to investigate the association between CST measurements and ORA measured corneal hysteresis (CH).
METHODS
Measurements of CST, ORA, axial length, average corneal curvature, central corneal thickness (CCT) and intraocular pressure with Goldmann applanation tonometry were carried out in 104 eyes of 104 patients with primary open-angle glaucoma and 35 eyes from normative subjects. The association between CST and ORA parameters was assessed using linear regression analysis, with model selection based on the second order bias corrected Akaike Information Criterion index.
RESULTS
Deformation amplitude ratio (corneal softness, R=-0.51), SP A1 (corneal stiffness, R=0.41), and Inverse Radius (integrated area under the curve of the inverse concave radius, R=-0.44) were significantly correlated with CH (P <0.05). The optimal model to explain CH using CST measurements was given by: CH=-76.3+4.6×A1 time (applanation time in the corneal inward movement)+1.9×A2 time (second applanation time in the corneal outward movement) + 3.1 × highest concavity deformation amplitude (magnitude of movement of the corneal apex from before deformation to its highest concavity) + 0.016×CCT (R=0.67; P<0.001).
CONCLUSIONS
CST parameters are significant, but weakly or moderately, related to ORA measured CH.
More by Takashi Fujishiro
View full profile →Structure Function Relationships of Vessel Density and Retinal Nerve Fiber Layer Thickness in Early Glaucomatous Eyes With High Myopia.
Contributing factors for intraocular pressure control in patients with mostly normal-tension glaucoma after initial Ex-PRESS drainage device implantation.
Top Research in Cornea & Biomechanics
Browse all →The impact of chronic use of prostaglandin analogues on the biomechanical properties of the cornea in patients with primary open-angle glaucoma.
Corneal Biomechanics and Visual Field Progression in Eyes with Seemingly Well-Controlled Intraocular Pressure.
Evaluation of a New Rebound Self-tonometer, Icare HOME: Comparison With Goldmann Applanation Tonometer.
Discussion
Comments and discussion will appear here in a future update.