A Novel Indentation Assessment to Measure Corneal Biomechanical Properties in Glaucoma and Ocular Hypertension.
Yunzhi Xu, Yiming Ye, Iok Tong Chong, Zidong Chen, Jiangang Xu, Yangfan Yang, Keming Yu, David C C Lam, Minbin Yu
Summary
Parameters derived from the CID and Corvis ST demonstrated agreement in the measurement of corneal biomechanical properties. The stiffness and modulus can characterize in vivo corneal biomechanical properties.
Abstract
PURPOSE
To evaluate the ability of the new in vivo corneal indentation device (CID) to measure corneal biomechanical properties.
METHODS AND RESULTS
In total, 186 eyes from 46 healthy subjects, 107 patients with primary open-angle glaucoma, and 33 patients with ocular hypertension were enrolled in a cross-sectional study. Measurements were performed using corneal visualization Scheimpflug technology (Corvis ST) and the CID. The deformation amplitude (DA), inward applanation time, inward applanation velocity (A1V), outward applanation time (A2T), outward applanation velocity (A2V), highest concavity time, DA ratio, max inverse radius (MIR), integrated radius, and stiffness parameter A1 were included as Corvis ST parameters, and stiffness and modulus were included as CID parameters. Associations between the Corvis ST and CID parameters and correlations between central corneal thickness and corneal biomechanical parameters were analyzed. The stiffness was significantly correlated with all the Corvis ST parameters (P < 0.05). The modulus was significantly correlated with the DA, A1V, A2T, A2V, highest concavity time, and MIR (P < 0.05). The DA, inward applanation time, A1V, A2T, A2V, DA ratio, MIR, integrated radius, and stiffness parameter A1 values and both CID-derived values were significantly correlated with central corneal thickness (P < 0.05).
CONCLUSIONS
Parameters derived from the CID and Corvis ST demonstrated agreement in the measurement of corneal biomechanical properties. The stiffness and modulus can characterize in vivo corneal biomechanical properties.
TRANSLATIONAL RELEVANCE
Agreeing with the Corvis ST regarding the assessment of corneal biomechanical properties, the CID can be a novel clinical tool for biomechanical evaluation of the cornea.
More by Yunzhi Xu
View full profile →Corneal Stiffness and Modulus of Normal-Tension Glaucoma in Chinese.
The Impact of Intraocular Pressure Changes on Corneal Biomechanics in Primary Open-angle Glaucoma.
Wide Corneal Epithelial Thickness Mapping in Eyes With Topical Antiglaucoma Therapy Using Optical Coherence Tomography.
Top Research in Cornea & Biomechanics
Browse all →The impact of chronic use of prostaglandin analogues on the biomechanical properties of the cornea in patients with primary open-angle glaucoma.
Corneal Biomechanics and Visual Field Progression in Eyes with Seemingly Well-Controlled Intraocular Pressure.
Evaluation of a New Rebound Self-tonometer, Icare HOME: Comparison With Goldmann Applanation Tonometer.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.