The effect of optic disc size on diagnostic precision with the Heidelberg retina tomograph.
Iester M, Mikelberg F S, Drance S M
AI Summary
This study found the HRT detected glaucoma well, but its diagnostic precision was lower for unusually small optic discs, highlighting a clinical challenge in these cases.
Abstract
Purpose
The authors evaluated the ability of a confocal scanning laser ophthalmoscope to detect glaucomatous visual field loss by using their previously described discriminant formula on a prospectively obtained cohort. The relationship of optic disc size to diagnostic classification was also evaluated.
Methods
One eye was chosen randomly from each of 153 subjects. Sixty control eyes had intraocular pressure less than 21 mmHg and normal visual fields; 93 glaucomatous eyes had intraocular pressure greater than 21 mmHg and abnormal visual fields. The optic disc status purposely was not used for classification purposes. All subjects were examined with the Heidelberg Retina Tomograph (HRT; Heidelberg Engineering GMBH, Heidelberg, Germany) and Humphrey Perimeter, program 30-2 (Humphrey Instruments, Inc., San Leandro, CA). Visual fields were considered abnormal by the authors' previously published criteria. The HRT classification used age, adjusted cup shape measure, rim volume, and height variation contour to classify the optic disc as normal or glaucomatous. Then the authors assessed the sensitivity, specificity, and diagnostic precision for the entire group, and for three subsets classified by disc area: disc area less than 2 mm2, between 2 and 3 mm2, and more than 3 mm2.
Results
The entire group had a sensitivity, specificity, and diagnostic precision of 74%, 88%, and 80%, respectively. The specificity was 83% when disc area was less than 2 mm2 and improved to 89% when disc area was more than 2 mm2. The sensitivity tended to improve from 65% to 79%, and to 83% if the disc area increased, but the difference was not statistically significant.
Conclusions
In a prospective cohort of patients, the HRT discriminant analysis formula was capable of detecting glaucomatous visual field loss with good precision. Unusually small optic discs continue to present diagnostic difficulties.
MeSH Terms
Shields Classification
Related Articles5
Evaluation of visual function and OCT parameters in ethambutol-induced optic neuropathy: a longitudinal study.
Cohort StudyPrimary Visual Pathway Changes in Individuals With Chronic Mild Traumatic Brain Injury.
Case-Control StudyEvaluating glaucoma in myopic eyes: Challenges and opportunities.
ReviewRetinal Perfusion and Injury in Sepsis and after Major Surgery.
Observational StudyComparison of structure-function correlation among IMO visual function analyser and Humphrey field analyser.
Cross-Sectional StudyIs this article assigned to the wrong chapter(s)? Let us know.