Long-term astigmatism after intraocular pressure sensor implantation and nonpenetrating glaucoma surgery: EYEMATE-SC trial.
Colya N Englisch, Karl T Boden, Peter Szurman, Kaweh Mansouri, H Burkhard Dick, Esther M Hoffmann, Marc J Mackert, Paul Schütte, Achim Langenbucher, André Messias, Philip Wakili
Summary
Despite its suprachoroidal localization, this study indicated that the miniaturized EYEMATE-SC IOP sensor did not negatively affect long-term astigmatism after combined implantation with NPGS.
Abstract
PURPOSE
To investigate long-term astigmatism after combined nonpenetrating glaucoma surgery (NPGS) and implantation of the first miniaturized suprachoroidal intraocular pressure (IOP) sensor EYEMATE-SC.
SETTING
The study was conducted in 5 medical centers in 2 different countries.
DESIGN
Retrospective multicenter clinical study.
METHODS
Astigmatism of patients instrumented with the EYEMATE-SC IOP sensor was assessed over a follow-up period of 3 years. Refraction and corrected distance visual acuity (CDVA) were obtained preoperatively, after 6 months, 1, 2, and 3 years. A canaloplasty-operated patient cohort served as control. Astigmatism was evaluated using 3-dimensional power vector analysis involving the spherical equivalent M, and the Jackson crossed cylinder projections J 0 and J 45 . Exclusion criteria included neovascular and angle-closure glaucoma, myopia, axial length outside 22 to 26 mm, other ocular diseases, prior glaucoma surgery, other ocular surgery within 6 months (cataract surgery within 3 months) before NPGS, serious generalized conditions, and other active medical head/neck implants.
RESULTS
Multivariate analysis indicated no changes in astigmatism along the observation period in both the EYEMATE-SC (n = 24) and the canaloplasty (n = 24) group ( P > .05 or nonsignificant after Bonferroni correction). Astigmatism was unchanged between the EYEMATE-SC and the canaloplasty group at all timepoints ( P > .05). CDVA did not change along the observation period of 3 years in each of both groups ( P > .05).
CONCLUSIONS
Despite its suprachoroidal localization, this study indicated that the miniaturized EYEMATE-SC IOP sensor did not negatively affect long-term astigmatism after combined implantation with NPGS.
More by Colya N Englisch
View full profile →Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.