Intraocular Pressure Measurements With the Tono-Vera: A New Handheld Rebound Tonometer.
Summary
There was a strong correlation and agreement in IOP measurements between calibrated Tono-Vera IOP and ORA IOPg.
Abstract
PRCIS
In a cross-sectional study comparing the novel Tono-Vera handheld rebound tonometer with the Ocular Response Analyzer, a strong agreement was found between the intraocular pressure measurements of the 2 devices.
PURPOSE
To compare intraocular pressure (IOP) measurements obtained by the new Tono-Vera (Reichert Technologies), a novel handheld rebound tonometer with a proprietary camera-based positioning system, with those from the Ocular Response Analyzer (ORA, Reichert Technologies) non-contact tonometer.
PATIENTS
Fifty-four eyes of normal individuals and open angle glaucoma subjects.
METHODS
All subjects had at least 6 reliable measurements with the Tono-Vera and 3 with the ORA, acquired by the same technician in the same visit. The sample was randomly split into calibration and validation samples. An ordinary least squares linear regression model generated a calibration formula for Tono-Vera readings using ORA's Goldmann-correlated IOP (IOPg) as a target. Bland-Altman plots evaluated the agreement between Tono-Vera calibrated IOP (IOPtv) and ORA IOPg in the validation sample.
MAIN OUTCOME MEASURES
Agreement between IOPtv and IOPg measurements.
RESULTS
Fifty glaucoma and 4 normal eyes were included, mean age 68.3 ± 9.4 years, mean central corneal thickness 535.0 ± 39.6 μm. In the validation sample, mean IOPtv was 14.6 ± 5.2 mmHg (range, 7.4-28.5 mmHg) versus 15.0 ± 5.3 mmHg (range, 6.6-28.6 mmHg) for ORA IOPg ( R ² = 0.782). Bland-Altman plots showed a mean difference of -0.4 ± 2.5 mmHg (95% limits of agreement: -5.4 to 4.5 mmHg). 58.6%, 79.3%, and 96.5% of IOPtv measurements fell within ± 2, ± 3, and ± 5 mmHg of ORA measurements. No fixed or proportional biases were observed ( P = 0.620).
CONCLUSION
There was a strong correlation and agreement in IOP measurements between calibrated Tono-Vera IOP and ORA IOPg.
Keywords
More by Luiz A F Beniz
View full profile →Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.