Comparative Study of Early Safety and Effectiveness Outcomes of the PreserFlo MicroShunt with and without an Intraluminal Suture Stent.
Jason Chami, Jeremy C K Tan, David Manning, George Kong, Jason Cheng, Hamish Dunn, Anna Galanopoulos, Brian Chua, Joanne Thai, Mark Hassall, Jamie E Craig, Mitchell Lawlor
Summary
The use of an intraluminal suture stent in PreserFlo MicroShunt surgery reduces the incidence of early hypotony without compromising surgical success.
Abstract
OBJECTIVE
To assess the early safety and effectiveness outcomes of the PreserFlo MicroShunt with and without an intraluminal suture stent.
DESIGN
Multicenter observational retrospective study using data from the Fight Glaucoma Blindness registry.
PARTICIPANTS
A total of 183 eyes in 172 patients, with a mean age of 73 ± 14 years, who underwent PreserFlo MicroShunt surgery with/without intraluminal stent suture placement, and with at least 6 months of follow-up. Eyes were divided into 2 groups: stent (68 eyes) and no-stent (115 eyes).
METHODS
Baseline and postoperative measurements of intraocular pressure (IOP), visual acuity (VA), number of glaucoma medications, and adverse events were recorded at 1, 3, and 6 months. Outcomes were compared between stent and no-stent groups using t tests for continuous variables and Fisher exact tests for categorical variables.
MAIN OUTCOME MEASURES
The primary outcome measures were the incidence of numerical hypotony (IOP ≤ 5 mmHg) and symptomatic hypotony (numerical hypotony plus ≥10 letters of VA loss) within 1 month, 1 to 3 months, and 3 to 6 months postoperatively. Secondary outcome measures included surgical success rates (defined as IOP ≤12, ≤15, ≤18, or ≤21 mmHg and ≥20% IOP reduction without hypotony and without additional glaucoma medications) and the need for secondary surgical interventions.
RESULTS
In the first postoperative month, the stent group had significantly lower rates of numerical hypotony (24% vs. 44%; P = 0.007) and symptomatic hypotony (13% vs. 28%; P = 0.027) than the no-stent group. The rate of device revision, explant, or replacement with an alternative shunt was also significantly lower in the stent group (3 eyes; 4.4%) than in the no-stent group (17 eyes; 14.8%; P = 0.047). No significant differences in hypotony rates were observed in the later follow-up windows (1-3 and 3-6 months), nor were there significant differences in surgical success rates between the stent and no-stent groups at any time point.
CONCLUSIONS
The use of an intraluminal suture stent in PreserFlo MicroShunt surgery reduces the incidence of early hypotony without compromising surgical success. These findings suggest that routine use of intraluminal sutures may improve early postoperative safety. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
Keywords
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.