Reproducibility and clinical evaluation of rebound tonometry.
Martinez-de-la-Casa Jose M, Garcia-Feijoo Julian, Castillo Alfredo, Garcia-Sanchez Julian
AI Summary
Rebound tonometry was studied for reproducibility and comparison with Goldmann applanation. It's reproducible, but consistently overestimates IOP compared to Goldmann, and both are similarly affected by corneal thickness.
Abstract
Purpose
To establish the reproducibility of a rebound tonometer in humans and the effect of corneal thickness on measurements, comparing it with Goldmann applanation tonometer.
Methods
In a first study designed to examine the reliability of the RBT, three experienced ophthalmologists undertook three consecutive intraocular pressure (IOP) measurements in 12 eyes of 12 normal subjects. A cross-sectional study was then performed to compare measurements obtained using the two tonometers in 147 eyes of 85 patients with ocular hypertension or glaucoma.
Results
Intraobserver coefficients of correlation obtained in the reproducibility study were 0.82, 0.73, and 0.87. Interobserver correlation was 0.82. There was a good correlation between IOP readings obtained by the RBT and the GAT (r = 0.865, P < 0.0001). RBT readings were consistently higher than GAT measurements (median difference, 1.8 +/- 2.8 mm Hg). A Bland-Altman plot indicated the 95% limits of agreement between the two methods were -3.7 to 7.3 mm Hg (slope = -0.022, P = 0.618). Using RBT, the point that best discriminated between patients with an IOP < or = 21 mm Hg and those with >21 mm Hg, as determined by the GAT was >23 mm Hg (sensitivity, 70.5%; specificity, 95.1%). In terms of pachymetry, the two tonometers behaved in a similar way, with correlation observed between IOP measurements and central corneal thickness.
Conclusions
Rebound tonometry is a reproducible method of determining IOP in humans. In general, it tends to overestimate IOP compared with Goldmann applanation tonometry. The tonometers used in both methods are similarly affected by pachymetry.
MeSH Terms
Shields Classification
Related Articles5
Development and Evaluation of an Artificial Intelligence Model to Set Target IOP for Glaucoma.
Observational StudyThe relationship between intraocular pressure and glaucoma: An evolving concept.
ReviewLong-term Safety and Performance of a Suprachoroidal Pressure Sensor System: Results of the EYEMATE-SC Trial Follow-up Study.
Clinical TrialGlaucoma Home Self-Testing Using VR Visual Fields and Rebound Tonometry Versus In-Clinic Perimetry and Goldmann Applanation Tonometry: A Pilot Study.
Clinical TrialContinuous 24-hour intraocular pressure monitoring in normal Chinese adults using a novel contact lens sensor system.
Observational StudyIs this article assigned to the wrong chapter(s)? Let us know.