Level of disagreement between Proview phosphene tonometer and Goldmann applanation tonometer intraocular pressure readings.
Rai Sushma, Moster Marlene R, Kesen Muge, Fontanarosa Joann, Spaeth George L, Steinmann William C, Wilson Richard, Myers Jonathan, Henderer Jeffrey
AI Summary
This study found significant disagreement between Proview and Goldmann IOP readings. However, regular Proview use significantly reduced glaucoma patients' anxiety, suggesting potential psychological benefits despite poor IOP accuracy.
Abstract
Purpose
To evaluate the agreement of intraocular pressure (IOP) readings obtained with the Proview phosphene tonometer and those obtained by Goldmann applanation tonometry as well as the effect of regular use of the Proview on patients' anxiety about their glaucoma.
Methods
One hundred thirty-five consecutive patients with glaucoma, 35 designated as controls, were enrolled in a 10-month randomized prospective clinical trial. The study patients but not the controls used the Proview outside the office. At office visits IOP was measured by an ophthalmologist with the Goldmann applanation tonometer (GAT) as well as by examiners and patients using the Proview. Patients in the treatment group (n = 100) were asked to measure and record their IOP at home as well. Using ANOVA, we examined the relationship between demographic variables and the level of agreement between the Proview readings and those measured by GAT at the final office visit. All participants also were asked to complete a questionnaire regarding their anxiety about their IOP at the baseline and final office visits. The primary outcome measure was the level of agreement of the Proview measurements with those obtained by GAT. Secondary outcome measures included patients' anxiety about their glaucoma.
Results
The absolute mean difference between GAT and Proview readings at the final visit was 3.5 +/- 2.9 mm Hg (median, 2.8 mm Hg). The treatment group reported significantly less anxiety about their glaucoma after 4 to 6 weeks of using the Proview (P = .024).
Conclusions
There was considerable discrepancy between Proview and GAT readings. However, regular use of the Proview tonometer significantly reduced patients' anxiety about their glaucoma.
MeSH Terms
Shields Classification
Related Articles5
The Paul Glaucoma Implant: a systematic review of safety, efficacy, and emerging applications.
Systematic ReviewRe-Esterified Triglyceride ω-3 Fatty Acids in Dry Eye Disease With Meibomian Gland Dysfunction: A Randomized Clinical Trial.
Randomized Controlled TrialLong-Term Outcomes of Iris-Fixated Phakic Intraocular Lenses in Low Myopia: A Retrospective Analysis With Over a Decade of Follow-up.
Case SeriesSelf-monitoring of the intraocular pressure with ICARE® HOME2 and a smartphone application: Feasibility and acceptability.
Clinical TrialTreatment Outcomes Comparing the Paul and Baerveldt Glaucoma Implants After One Year of Follow-Up.
Retrospective StudiesIs this article assigned to the wrong chapter(s)? Let us know.