Reproducibility and clinical relevance of the ocular response analyzer in nonoperated eyes: corneal biomechanical and tonometric implications.
Moreno-Montañés Javier, Maldonado Miguel J, García Noelia, Mendiluce Loreto, García-Gómez Pio J, Seguí-Gómez María
AI Summary
The ORA provides reproducible corneal biomechanical and IOP measurements. Corneal biomechanics significantly adjust IOP readings in over a quarter of eyes, impacting glaucoma assessment.
Abstract
Purpose
To assess the reproducibility of the ocular response analyzer (ORA) in nonoperated eyes and the impact of corneal biomechanical properties on intraocular pressure (IOP) measurements in normal and glaucomatous eyes.
Methods
In the reliability study, two independent examiners obtained repeated ORA measurements in 30 eyes. In the clinical study, the examiners analyzed ORA and IOP-Goldmann values from 220 normal and 42 glaucomatous eyes. In both studies, Goldmann-correlated IOP measurement (IOP-ORAg), corneal-compensated IOP (IOP-ORAc), corneal hysteresis (CH), and corneal resistance factor (CRF) were evaluated. IOP differences of 3 mm Hg or greater between the IOP-ORAc and IOP-ORAg were considered outcome significant.
Results
Intraexaminer intraclass correlation coefficients and interexaminer concordance correlation coefficients ranged from 0.78 to 0.93 and from 0.81 to 0.93, respectively, for all parameters. CH reproducibility was highest, and the IOP-ORAg readings were lowest. The median IOP was 16 mm Hg with the Goldmann tonometer, 14.5 mm Hg with IOP-ORAg (P < 0.001), and 15.7 mm Hg with IOP-ORAc (P < 0.001). Outcome-significant results were found in 77 eyes (29.38%). The IOP-ORAc, CH, and CRF were correlated with age (r = 0.22, P = 0.001; r = -0.23, P = 0.001; r = -0.14, P = 0.02, respectively), but not the IOP-ORAg or IOP-Goldmann.
Conclusions
The ORA provides reproducible corneal biomechanical and IOP measurements in nonoperated eyes. Considering the effect of ORA, corneal biomechanical metrics produces an outcome-significant IOP adjustment in at least one quarter of glaucomatous and normal eyes undergoing noncontact tonometry. Corneal viscoelasticity (CH) and resistance (CRF) appear to decrease minimally with increasing age in healthy adults.
MeSH Terms
Shields Classification
Related Articles5
Development and Evaluation of an Artificial Intelligence Model to Set Target IOP for Glaucoma.
Observational StudyThe relationship between intraocular pressure and glaucoma: An evolving concept.
ReviewLong-term Safety and Performance of a Suprachoroidal Pressure Sensor System: Results of the EYEMATE-SC Trial Follow-up Study.
Clinical TrialGlaucoma Home Self-Testing Using VR Visual Fields and Rebound Tonometry Versus In-Clinic Perimetry and Goldmann Applanation Tonometry: A Pilot Study.
Clinical TrialContinuous 24-hour intraocular pressure monitoring in normal Chinese adults using a novel contact lens sensor system.
Observational StudyIs this article assigned to the wrong chapter(s)? Let us know.