Improved Outcomes for Transscleral Cyclophotocoagulation Through Optimized Treatment Parameters.

Quigley Harry A


AI Summary

This study found transscleral cyclophotocoagulation for glaucoma is more successful when power is tailored to audible cues and higher total energy is delivered, rather than using standardized protocols.

Abstract

Purpose

To compare outcomes of transscleral diode cyclophotocoagulation with the treatment parameters used.

Patients and methods: This was a retrospective chart review of a random, 50% sample of diode procedures using the G-probe over 10 years for uncontrolled glaucoma. The main outcome measure was intraocular pressure (IOP) reduction by 20% and final IOP≤21 mm Hg.

Results

In 236 eyes (persons) treated by 5 glaucoma specialists, most eyes had severe glaucoma, with 75% having <20/200 acuity. Median follow-up was 2.7 years. In eyes receiving only 1 treatment, IOP success criterion was met in 72% (129/180). Success was significantly related to power per delivery and median total joules per treatment (successes=135 J, failures=98 J; P=0.0009), but not to number of deliveries, nor to extent of circumference treated. Greater success was associated with 3 or 4 second duration per delivery, power level based on audible cues, and firm pressure on the sclera. Using a standard 2000 mW, 2 seconds, 20 deliveries in each eye had the lowest success (49%). Of those with no preoperative pain, 40 persons (57%) had no postoperative pain, whereas 20 reported pain of 1-3/10 (29%). Phthisis occurred in 7 eyes (3%), 5 of which had severe secondary eye disease. Nine eyes had no light perception preoperatively, whereas 50 eyes were no light perception at last follow-up, many after additional surgeries for other conditions.

Conclusions

Diode cyclophotocoagulation achieved reasonable IOP lowering, often without severe postoperative pain or complication. Greater success was achieved when audible effects were used to tailor the power settings to individual responses. Diode treatments with no intraoperative effect adjustment or using standardized protocols may not achieve optimal success.


MeSH Terms

AdultAgedAged, 80 and overCiliary BodyFemaleGlaucoma, Open-AngleHumansIntraocular PressureLaser CoagulationLasers, SemiconductorMaleMiddle AgedOcular HypotensionRetrospective StudiesScleraTreatment OutcomeVisual Acuity

Key Concepts5

Transscleral diode cyclophotocoagulation achieved an intraocular pressure (IOP) success criterion (20% reduction and final IOP ≤ 21 mm Hg) in 72% (129/180) of eyes receiving only 1 treatment.

TreatmentCase seriesRetrospective chart reviewn=180 eyes (from 236 eyes total) receiv…Ch44

The success of transscleral diode cyclophotocoagulation was significantly related to power per delivery and median total joules per treatment (successes=135 J, failures=98 J; P=0.0009), but not to the number of deliveries or the extent of circumference treated.

TreatmentCase seriesRetrospective chart reviewn=236 eyesCh44

Greater success with transscleral diode cyclophotocoagulation was associated with 3 or 4 second duration per delivery, power level based on audible cues, and firm pressure on the sclera.

TreatmentCase seriesRetrospective chart reviewn=236 eyesCh44

Using a standard 2000 mW, 2 seconds, 20 deliveries protocol for transscleral diode cyclophotocoagulation resulted in the lowest success rate (49%).

TreatmentCase seriesRetrospective chart reviewn=236 eyesCh44

Phthisis occurred in 7 eyes (3%) after transscleral diode cyclophotocoagulation, with 5 of these eyes having severe secondary eye disease.

PrognosisCase seriesRetrospective chart reviewn=236 eyesCh44