Cyclocoagulation of the ciliary bodies by high-intensity focused ultrasound: a 12-month multicenter study.
Philippe Denis, Florent Aptel, Jean-François Rouland, Jean-Philippe Nordmann, Yves Lachkar, Jean-Paul Renard, Eric Sellem, Christophe Baudouin, Alain Bron
Summary
The new miniaturized HIFU EyeOP1 delivery device seems to be effective in decreasing IOP in patients with refractory glaucoma. The technology offers a good safety profile. (ClinicalTrials.gov number, NCT01338467.).
Abstract
PURPOSE
To evaluate the safety and efficacy of high-intensity focused ultrasound (HIFU) cyclocoagulation in reducing intraocular pressure (IOP) in patients with refractory glaucoma by using a novel miniaturized delivery device (EyeOP1).
METHODS
We conducted a 12-month open-label multicenter prospective study (EyeMUST1 Study). Patients with primary (primary open-angle glaucoma [POAG]) or secondary refractory glaucoma were treated in two groups depending on the duration of each ultrasound shot (group 1: 4 seconds; group 2: 6 seconds). The primary efficacy outcome was based on IOP reduction at 6 and 12 months.
RESULTS
Fifty-two patients were enrolled: 36 (69%) had POAG and 16 (31%) had secondary glaucoma. Group 1 (n = 24) and group 2 (n = 28) had similar demographics and baseline characteristics. In group 1, IOP was reduced from a mean preoperative value of 29.7 ± 7.7 mm Hg (n = 3.5 glaucoma medications) to a mean postoperative value of 21.3 ± 6.7 mm Hg (n = 3.5 glaucoma medications) and 20.1 ± 6.7 mm Hg (n = 3.2 glaucoma medications) at 6 and 12 months, respectively. In group 2, IOP was reduced from a mean preoperative value of 29.0 ± 7.4 mm Hg (n = 3.3 glaucoma medications) to a mean postoperative value of 20.2 ± 7.4 mm Hg (n = 3.4 glaucoma medications) and 18.5 ± 6.6 mm Hg (n = 3.5 glaucoma medications) at 6 and 12 months, respectively. At 12 months, the IOP reduction was sustained in both groups (32% IOP reduction in group 1 and 36% IOP reduction in group 2). The overall tolerance of the technique was good, with no serious adverse events.
CONCLUSIONS
The new miniaturized HIFU EyeOP1 delivery device seems to be effective in decreasing IOP in patients with refractory glaucoma. The technology offers a good safety profile. (ClinicalTrials.gov number, NCT01338467.).
Keywords
More by Philippe Denis
View full profile →A First-in-Human Study of the Efficacy and Safety of MINIject in Patients with Medically Uncontrolled Open-Angle Glaucoma (STAR-I).
A European Study of the Performance and Safety of MINIject in Patients With Medically Uncontrolled Open-angle Glaucoma (STAR-II).
Two-Year Performance and Safety Results of the MINIject Supraciliary Implant in Patients With Primary Open-Angle Glaucoma: Meta-Analysis of the STAR-I, II, III Trials.
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Discussion
Comments and discussion will appear here in a future update.