Macular ganglion cell layer imaging in preperimetric glaucoma with speckle noise-reduced spectral domain optical coherence tomography.
Nakano Noriko, Hangai Masanori, Nakanishi Hideo, Mori Satoshi, Nukada Masayuki, Kotera Yuriko, Ikeda Hanako Ohashi, Nakamura Hajime, Nonaka Atsushi, Yoshimura Nagahisa
AI Summary
Speckle noise-reduced SD-OCT clearly showed macular GCL thinning in preperimetric glaucoma, especially inferiorly. This method offers high sensitivity for early glaucoma detection, improving diagnosis before vision loss.
Abstract
Objective
To visualize the macular ganglion cell layer (GCL) and measure its thickness in normal eyes and eyes with preperimetric glaucoma, using speckle noise-reduced spectral domain optical coherence tomography (SD-OCT).
Design
Retrospective consecutive case series.
Participants
Thirty-seven eyes of 37 patients with preperimetric glaucoma and 39 normal eyes of 39 volunteers.
Methods
Vertical and horizontal SD-OCT B-scan images were acquired with minimal speckle noise by using eye-tracking to obtain and average 50 B-scans at each identical location of interest. B-scan images were manually analyzed for GCL, retinal nerve fiber layer (RNFL), and inner plexiform layer shapes and thicknesses in the macula.
Main outcome measures
Macular GCL images and thickness in normal eyes and in eyes with preperimetric glaucoma.
Results
The macular GCL was clearly seen on speckle noise-reduced SD-OCT images in normal eyes and eyes with preperimetric glaucoma. In each eye with preperimetric glaucoma, thinning of the macular GCL was visually apparent, particularly on vertical scans. The mean regional macular GCL was most severely thinned in the inferior perifoveal region, where its thickness was <70% of its normal thickness in 30 (81.1%) of the 37 eyes and <50% of its normal thickness in 13 (35.1%) of the 37 eyes. When the sensitivity and specificity for detecting abnormal thinning (outside the lower limit of 99% confidence interval [CI] for the means in the 39 normal eyes) in at least one 0.5-mm segment or sector were compared, the macular GCL on vertical B-scans exhibited higher sensitivity (81.1%) than the other layers on vertical B-scans (99% CI, 5.4%-59.5%; P = 0.00075-0.02100), the macular GCL (99% CI, 40.5%; P = 0.00027) on horizontal B-scans, the other layers (99% CI, 5.4%-48.6%; P<0.00048-0.00400) on horizontal B-scans, and circumpapillary RNFL automatically measured on SD-OCT (54.1%; P = 0.021), and scanning laser polarimetry with variable corneal compensation (24.3%; P = 0.00095). All the macular layers on both the vertical and horizontal B-scans and circumpapillary RNFL thickness exhibited comparable specificity (91.4-100.0%, statistically not different).
Conclusions
Speckle noise-reduced SD-OCT imaging allowed clear visualization and measurement of the macular GCL, which was severely thinned in eyes with preperimetric glaucoma.
Financial disclosure(s): Proprietary or commercial disclosure may be found after the references.
MeSH Terms
Shields Classification
Related Articles5
Prediction of Visual Field Progression with Baseline and Longitudinal Structural Measurements Using Deep Learning.
Observational StudyAssessing Glaucoma Severity and Progression in Individuals with Asymmetric Axial Length: An Intrapatient Comparative Study.
Observational StudyLong-Term Rate of Optic Disc Rim Loss in Glaucoma Patients Measured From Optic Disc Photographs With a Deep Neural Network.
Cohort StudyPeripapillary Atrophy Area as an Indicator of Glaucomatous Structural and Functional Progression.
Cohort StudyDetecting Visual Field Worsening From Optic Nerve Head and Macular Optical Coherence Tomography Thickness Measurements.
Observational StudyIs this article assigned to the wrong chapter(s)? Let us know.