Global Search

Search articles, concepts, and chapters

Br J OphthalmolNovember 201378 citations

Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma.

Reznicek Lukas, Muth Daniel, Kampik Anselm, Neubauer Aljoscha S, Hirneiss Christoph


AI Summary

This study found a novel Scheimpflug device (Corvis ST) accurately measures IOP and CCT in healthy, OHT, and glaucoma patients, showing good agreement with standard methods and excellent repeatability.

Abstract

Background

To evaluate the agreement of intraocular pressure (IOP) and central corneal thickness (CCT) measurements obtained with the non-contact tonometer Corvis Scheimpflug Technology (Corvis ST, OCULUS, Wetzlar, Germany) versus Goldmann applanation tonometry (GAT) and ultrasound-based corneal pachymetry (US-CCT).

Methods

Eye healthy participants, patients with ocular hypertension (OHT) and patients with open-angle glaucoma were included in this prospective study. In each participant, GAT, US-CCT and measurements with Corvis ST were obtained (Corvis-IOP and Corvis-CCT). Accuracy and repeatability were tested by correlation and regression analyses, Bland-Altman plots and assessment of intraclass correlation coefficients.

Results

A consecutive series of 188 right study eyes of 188 participants (142 eyes with glaucoma, 10 eyes with OHT and 36 control eyes) were included in this prospective study. The mean GAT of all included was 14.5±4.8 mm Hg compared with mean Corvis-IOP of 15.4±5.6 mm Hg (Spearman's r=0.75, p<0.0001). Mean US-CCT was 544.56±40.0 µm compared with Corvis-CCT of 545.2±46.5 µm (Pearson's r=0.78, p<0.0001). Bland-Altman plots of all included eyes as well as subgroup analyses revealed good agreement of the IOP and CCT measurement techniques. High intraclass correlation coefficient values in 17 patients with repeated measurements revealed very good repeatability (0.942 and 0.937 for Corvis-IOP and Corvis-CCT, respectively). Corvis-IOP but not GAT showed a trend of dependence on CCT.

Conclusions

Obtaining CCT and measuring IOP with the Corvis ST reveals very good repeatability and good accuracy in healthy subjects and patients with OHT and glaucoma when compared with standardised US pachymetry or GAT.


MeSH Terms

CorneaCorneal PachymetryFemaleFollow-Up StudiesGlaucoma, Open-AngleHumansIntraocular PressureMaleMiddle AgedOcular HypertensionProspective StudiesReproducibility of ResultsTonometry, Ocular

Is this article assigned to the wrong chapter(s)? Let us know.