Comparison of the Intraocular Pressure Measured Using the New Rebound Tonometer Icare ic100 and Icare TA01i or Goldmann Applanation Tonometer.
Shunsuke Nakakura, Etsuko Mori, Yuki Fujio, Yasuko Fujisawa, Kanae Matsuya, Yui Kobayashi, Hitoshi Tabuchi, Ryo Asaoka, Yoshiaki Kiuchi
Summary
The new rebound tonometer Icare ic100 almost constantly showed IOPs lower than GAT. The difference was affected by CCT, corneal curvature, and disease.
Abstract
PURPOSE
We evaluated the agreement between the intraocular pressure (IOP) values of new rebound tonometer, Icare ic100, and Icare TA01i or Goldmann applanation tonometer (GAT).
METHODS
We studied one eye each of 106 subjects (57 with primary open-angle glaucoma, 49 healthy subjects). IOP was randomly measured twice with the patient in sitting position using the Icare ic100, Icare TA01i, and GAT. Tonometer measurements were evaluated using Bland-Altman analysis. The relationship between IOP difference (Icare ic100-GAT) and age, sex, disease, axial length, central corneal thickness (CCT), and corneal curvature was investigated using multivariate regression analysis.
RESULTS
IOPs measured using Icare ic100, Icare TA01i, and GAT were 11.7±3.0 (mean±standard deviation), 12.2±2.9, and 16.0±3.2 mm Hg, respectively (P0.05; Tukey-Kramer test). Bland-Altman analysis revealed that the mean differences between Icare ic100 and Icare TA01i and those between Icare ic100 and GAT were -0.46 and -4.2 mm Hg, respectively (95% limits of agreement, -3.35 to 2.42 and -10.10 to 1.61 mm Hg, respectively). For IOP differences between Icare ic100 and GAT, parameters selected in the optimal model were CCT (coefficient, 20.3, P=0.029), corneal curvature (3.0, P=0.020), and glaucoma-normal (-1.0, P=0.004).
CONCLUSIONS
The new rebound tonometer Icare ic100 almost constantly showed IOPs lower than GAT. The difference was affected by CCT, corneal curvature, and disease.
More by Shunsuke Nakakura
View full profile →Deep-learning Classifier With an Ultrawide-field Scanning Laser Ophthalmoscope Detects Glaucoma Visual Field Severity.
The Relationship Between Corvis ST Tonometry Parameters and Ocular Response Analyzer Corneal Hysteresis.
Association between Corneal Biomechanical Properties with Ocular Response Analyzer and Also CorvisST Tonometry, and Glaucomatous Visual Field Severity.
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
Discussion
Comments and discussion will appear here in a future update.