Assessing Agreement of Intraocular Pressure from Four Mechanistically Different Tonometers in the Eye Dynamics and Engineering Network Clinical Trial.
Max Liu, David M Reed, Shan Fan, Arash Kazemi, Vikas Gulati, Arthur J Sit, Sayoko E Moroi, Carol B Toris
Summary
When adjusted for proparacaine's effect on IOP, 3 tonometers were equivalent between each other when determined by equivalence statistical testing with equivalence bounds ± 2 mmHg.
Abstract
PURPOSE
To compare the measurements of intraocular pressure (IOP) by four mechanistically different tonometers in healthy participants from the Eye Dynamics and Engineering Network Aqueous Humor Dynamics clinical trial using the mean difference estimation, equivalence statistics, and adjusting for proparacaine's effects on IOP.
DESIGN
Multicenter open-label clinical trial in healthy adults (NCT01677507).
PARTICIPANTS
Enrolled were 122 participants (55.2 ± 8.8 years; 92 women and 30 men), 121 of which had measurements with and without proparacaine.
METHODS
Participants from 3 sites had their IOPs measured with iCare rebound tonometer (RT), ocular response analyzer corneal-compensated (CC), pneumatonometer (PN), and Goldmann applanation tonometer (GAT). Intraocular pressure readings with RT and CC were collected with and without proparacaine.
MAIN OUTCOME MEASURES
Effect sizes and agreement in IOP measurements from the four tonometers was analyzed with Data analysis using Bootstrap-Coupled ESTimation, Bland-Altman Limits of Agreement (LoAs), Lin's concordance correlation coefficient (CCC), and robust equivalence two 1-sided t-tests in tandem with null hypothesis significance testing. Data analysis using bootstrap-Coupled ESTimation was used to analyze differences in IOP readings from RT and CC before and after administering proparacaine.
RESULTS
Data analysis using Bootstrap-Coupled ESTimation, LoA, and CCC showed greatest effect size and poorest agreement between PN and the other tonometers. The lowest mean difference was seen when comparing GAT and CC. However, LoA and CCC found the strongest agreement between RT and GAT. Two 1-sided tests revealed tonometers can be considered equivalent to one another at ± 2 mmHg, except for PN. After applying proparacaine, IOP decreased significantly (P 2 mmHg higher than RT, CC, or GAT. Therefore, when comparing IOP data from different tonometers, one should consider that not all tonometer measurements are interchangeable and the proparacaine effect should be considered. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
Keywords
Top Research in IOP & Medical Therapy
Browse all →The Complications of Myopia: A Review and Meta-Analysis.
Inflammation in Glaucoma: From the back to the front of the eye, and beyond.
Treatment Outcomes in the Primary Tube Versus Trabeculectomy Study after 1 Year of Follow-up.
In the Knowledge Library
Discussion
Comments and discussion will appear here in a future update.