Global Search

Search articles, concepts, and chapters

Graefes Arch Clin Exp OphthalmolFebruary 20062 citations

Repeatability and reproducibility of optic nerve head topography using the retinal thickness analyzer.

Hoffmann Esther M, Medeiros Felipe A, Kramann Christina, Pfeiffer Norbert, Grus Franz H


AI Summary

This study found RTA measurements of optic nerve topography have high variability, making them unreliable for precise glaucoma diagnosis and monitoring.

Abstract

Objective

The objective was to determine the reproducibility and repeatability of topographic disk parameters with the retinal thickness analyzer (RTA).

Methods

Thirty-nine eyes of 39 healthy subjects were included. Three observers performed three consecutive measurements of optic disk topography independently. Combined repeatability (intra-observer component of variability) and reproducibility (inter-observer component of variability, R and R) was assessed by variance component analysis.

Results

Six topographic parameters (disk area, cup area, cup volume, mean cup depth, maximum depth, and cup shape measure) revealed an acceptable repeatability and reproducibility expressed as the percentage of the total measurement variability (range 17.6-28.4%). The average reproducibility and repeatability of all optic disk parameters measured using the RTA was not acceptable (combined R and R = 33.4%).

Conclusions

Repeatability and reproducibility of the optic disk topography analysis using the RTA reveals a high variation, and should, therefore, be interpreted carefully for glaucoma diagnosis and for monitoring the disease.


MeSH Terms

AdultDiagnostic Techniques, OphthalmologicalFemaleHumansIntraocular PressureMaleMiddle AgedObserver VariationOptic DiskReproducibility of ResultsRetinaVisual Acuity

Is this article assigned to the wrong chapter(s)? Let us know.