Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography a study on diagnostic agreement with Heidelberg Retinal Tomograph.
Leung Christopher Kai-shun, Ye Cong, Weinreb Robert N, Cheung Carol Yim Lui, Qiu Quanliang, Liu Shu, Xu Guihua, Lam Dennis Shun Chiu
AI Summary
This study found Spectralis OCT retinal nerve fiber layer imaging had higher sensitivity than HRT optic disc measurements for glaucoma detection, suggesting superior diagnostic performance.
Abstract
Objective
To evaluate and compare the diagnostic agreement and performance for glaucoma detection between a confocal scanning laser ophthalmoscope and a spectral-domain optical coherence tomograph (OCT).
Design
Prospective, cross-sectional study.
Participants
One hundred fifty-five subjects (79 glaucoma and 76 normal subjects).
Methods
One eye from each individual was selected randomly for optic disc and retinal nerve fiber layer (RNFL) imaging by the Heidelberg Retinal Tomograph (HRT; Heidelberg Engineering, GmbH, Dossenheim, Germany) and the Spectralis OCT (Heidelberg Engineering), respectively. Glaucoma was defined based on the presence of visual field defects with the Humphrey visual field analyzer (Carl Zeiss Meditec, Dublin, CA). The agreement of the categorical classification ("within normal limits," "borderline," and "outside normal limits") at the temporal, superotemporal, superonasal, nasal, inferonasal and inferotemporal sectors of the optic disc were evaluated (kappa statistics). The diagnostic sensitivity and specificity between optic disc and RNFL assessment were compared (McNemar's statistics). Area under the receiver operating characteristic curve (AUC) of OCT RNFL and HRT optic disc parameters were computed after adjustment of age, axial length, and optic disc area.
Main outcome measures
Agreement of categorical classification, AUC of optic disc, and RNFL parameters.
Results
The agreement of categorical classification between HRT and Spectralis OCT were fair to moderate (kappa ranged between 0.30 and 0.53) except for global (kappa = 0.63) and inferotemporal (kappa = 0.68) measurements. Defining glaucoma as having "outside normal limits" in the global and/or in >or=1 of the sectoral measurements, the respective sensitivities of Spectralis OCT and HRT were 91.1% and 79.8% (P = 0.012) at a similar level of specificity (97.4% and 94.7%). The AUC of OCT global RNFL thickness (0.978) was greater than those of HRT global rim area (0.905), vertical cup-disc ratio (0.857), rim-disc area ratio (0.897), and multivariate discriminant analysis (0.880-0.925; all with P<or=0.028) after covariates adjustment.
Conclusions
The diagnostic classification provided in the HRT and Spectralis OCT analysis report may not agree. At a comparable level of specificity, Spectralis OCT RNFL measurement attained a higher sensitivity than HRT optic disc measurement.
MeSH Terms
Shields Classification
Related Articles5
Utilization of Image-Based Deep Learning in Multimodal Glaucoma Detection Neural Network from a Primary Patient Cohort.
Observational StudyA Deep Learning Model for Glaucoma Detection Outperforms Conventional Metrics.
Observational StudyDiagnostic Performance of a Handheld Electroretinography Test for Glaucoma.
Cross-Sectional StudyNormative Percentiles of Retinal Nerve Fiber Layer Thickness and Glaucomatous Visual Field Loss.
Observational StudyDetecting Glaucoma from Fundus Photographs Using Deep Learning without Convolutions: Transformer for Improved Generalization.
Evaluation of a diagnostic technologyIs this article assigned to the wrong chapter(s)? Let us know.