Invest Ophthalmol Vis SciSeptember 201145 citations

Comparison of standard automated perimetry, frequency-doubling technology perimetry, and short-wavelength automated perimetry for detection of glaucoma.

Liu Shu, Lam Shi, Weinreb Robert N, Ye Cong, Cheung Carol Y, Lai Gilda, Lam Dennis Shun-Chiu, Leung Christopher Kai-Shun


AI Summary

This study compared SAP, FDT, and SWAP for glaucoma detection, finding FDT and SAP performed similarly and better than SWAP, especially for early disease, offering guidance for test selection.

Abstract

Purpose

To compare the performance of standard automated perimetry (SAP), frequency-doubling technology (FDT) perimetry, and short-wavelength automated perimetry (SWAP) in detecting glaucoma.

Methods

One hundred thirty-two eyes of 95 glaucoma patients and 37 normal subjects had retinal nerve fiber layer (RNFL) imaging and visual field testing by SAP, Matrix FDT perimetry, and Swedish interactive thresholding algorithm (SITA) SWAP at the same visit (all perimeters by Carl Zeiss Meditec, Inc., Dublin, CA). Visual field defects were confirmed with two or more consecutive examinations by the same types of perimetry. Glaucoma was defined with the reference to the RNFL thickness deviation map score (≥ 4, glaucomatous; ≤ 2, normal). The sensitivity, specificity, and area under the receiver operating characteristic curve (AUC) of MD (mean deviation) and PSD (pattern standard deviation) of the perimetries were compared.

Results

Taking all glaucoma patients into consideration, the sensitivity was highest for Matrix FDT perimetry (69%), followed by SAP (68%), and then SITA SWAP (59%). When the analysis included only patients with early glaucoma, the sensitivity decreased to 52%, 46%, and 34%, respectively, with a significant difference detected between Matrix FDT perimetry and SITA SWAP (P = 0.034). The specificity was ≥ 97% for all perimetries. The AUCs of MD and PSD followed a similar order, with Matrix FDT perimetry having the greatest AUC (0.89-0.94), followed by SAP (0.87-0.94), and then SITA SWAP (0.69-0.90). There were significant differences in sensitivities at 90% specificity between Matrix FDT perimetry and SITA SWAP (P ≤ 0.005 for MD; P ≤ 0.039 for PSD).

Conclusions

The performance for glaucoma detection was comparable between FDT perimetry and SAP. FDT perimetry had a higher sensitivity for detecting glaucoma than did SWAP at a comparable level of specificity.


MeSH Terms

AlgorithmsArea Under CurveFalse Positive ReactionsGlaucomaHumansMiddle AgedNerve FibersPredictive Value of TestsROC CurveReproducibility of ResultsRetinal Ganglion CellsSensitivity and SpecificityTomography, Optical CoherenceVision DisordersVisual Field TestsVisual Fields