Comparison of a remote operating slit-lamp microscope system with a conventional slit-lamp microscope system for examination of trabeculectomy eyes.
Kashiwagi Kenji, Tanabe Naohiko, Go Kentaro, Imasawa Mitsuhiro, Mabuchi Fumihiko, Chiba Tetsuya, Sugiyama Atsushi, Abe Keitetsu
AI Summary
This study found remote slit-lamp systems can effectively examine trabeculectomy eyes, similar to conventional methods, but take longer, while 2D photography was less accurate.
Abstract
Purpose
To compare 2-dimensional (2D) photo imaging with slit-lamp photo imaging for examination of trabeculectomy eyes, and to compare the accuracy and consistency of examination of trabeculectomy eyes with a remote operating slit-lamp microscope system (referred to as "remote slit lamp" hereafter) and a conventional slit-lamp microscope system (referred to as "slit lamp" hereafter).
Methods
Thirty-five eyes of 35 patients having a history of trabeculectomy were enrolled in the study that compared 2D photo imaging with slit-lamp photo imaging for the evaluation of trabeculectomy eyes. Five ophthalmologists evaluated the 2D images and the slit-lamp images independently with masking of patient information. Evaluation scores were compared with those provided by a glaucoma specialist using the slit lamp as standard. Fifteen eyes from 15 patients having a history of trabeculectomy were enrolled in the study that investigated the accuracy and consistency of examination of trabeculectomy eyes with a remote slit lamp and a slit lamp.
Results
Central anterior chamber depth, bleb height, and bleb extent evaluated by the slit-lamp photo imaging showed significantly higher consistency with the standard than those evaluated by 2D photo imaging (P<0.05). The remote slit lamp showed good consistency of all the evaluated parameters with the slit lamp and the κ scores of all the evaluated parameters were higher than 0.8. The completion time for evaluation with the remote slit lamp and the slit lamp were 247.3±153.5 and 123.5±53.7 seconds, respectively (P<0.001).
Conclusions
Slit-lamp photo imaging is a superior method for examination of trabeculectomy eyes compared with 2D photo imaging. The remote slit-lamp system shows similar potential to the slit-lamp system for the evaluation of trabeculectomy eyes, although the evaluation time is much longer.
MeSH Terms
Shields Classification
Related Articles5
Anterior Segment Optical Coherence Tomography and Anterior Segment Optical Coherence Tomography Angiography After Bleb Forming Glaucoma Surgeries: A Systematic Review.
Systematic ReviewNovel Bleb Classification Following Ab Interno Implantation of Gel-Stent Using Anterior Segment Optical Coherence Tomography.
Cross-Sectional StudyStructural Confirmation of Lymphatic Outflow from Subconjunctival Blebs of Live Human Subjects.
Cross-Sectional StudyAnterior-segment optical coherence tomography of filtering blebs in the early postoperative period of ab externo SIBS microshunt implantation with mitomycin C: Morphological analysis and correlation with intraocular pressure reduction.
Observational StudyUpdates on the utility of anterior segment optical coherence tomography in the assessment of filtration blebs after glaucoma surgery.
ReviewIs this article assigned to the wrong chapter(s)? Let us know.