Visual function-specific perimetry to identify glaucomatous visual loss using three different definitions of visual field abnormality.
Tafreshi Ali, Sample Pamela A, Liebmann Jeffrey M, Girkin Christopher A, Zangwill Linda M, Weinreb Robert N, Lalezary Maziar, Racette Lyne
AI Summary
This study compared SAP, SWAP, and FDT perimetry for glaucoma detection. Finding: confirming visual field abnormality is important and best achieved with a repeat SAP or SWAP test.
Abstract
Purpose
To compare the most recent versions of standard automated perimetry (SAP), short-wavelength automated perimetry (SWAP), and frequency-doubling technology (FDT) using three definitions of visual field (VF) abnormality: single-test abnormality, abnormality confirmed by the same test, and abnormality confirmed by a different test.
Methods
Data obtained from one eye of each of 174 patients with glaucoma and 164 age-matched healthy control subjects from the Diagnostic Innovations in Glaucoma Study and African Descent and Glaucoma Evaluation Study were included, based on the appearance of the optic disc on stereophotographs. Each participant had two reliable 24-2 SAP-SITA, SWAP-SITA, and Matrix FDT tests. Receiver operating characteristic (ROC) curves were generated for the PSD of each test to equate the tests at 90% and 95% specificity. SAP, SWAP, and FDT were compared under each definition of VF abnormality by assessing the sensitivities, the agreement between tests and the overlap in deficit location at these set specificities. The tests were also compared using the machine-derived PSD.
Results
At a set specificity of 95%, single-test sensitivities of 30% (SAP), 29% (SWAP), and 28% (FDT) were observed (all P > 0.05). Sensitivities ranged from 24% to 27% (all P > 0.05) when same-test confirmation was used and from 20% to 23% (all P > 0.05) when different-test confirmation was used. SAP/SAP sensitivity was higher than all different-test combinations (all P < 0.05), and SWAP/FDT sensitivity was lower than all same-test combinations (all P < 0.05).
Conclusions
Confirming VF abnormality is important and optimal when an abnormal SAP is confirmed by a subsequent SAP or SWAP test.
MeSH Terms
Shields Classification
Related Articles5
Evaluation of visual function and OCT parameters in ethambutol-induced optic neuropathy: a longitudinal study.
Cohort StudyPrimary Visual Pathway Changes in Individuals With Chronic Mild Traumatic Brain Injury.
Case-Control StudyStructure-Function Associations Between Quantitative Contrast Sensitivity Function And Peripapillary Optical Coherence Tomography Angiography in Diabetic Retinopathy.
Cross-Sectional StudyRelating Standardized Automated Perimetry Performed With Stimulus Sizes III and V in Eyes With Field Loss Due to Glaucoma and NAION.
Observational StudyHigh-Resolution Microperimetry for Detecting Glaucomatous Damage: A Prospective Evaluation of Performance.
Prospective StudiesIs this article assigned to the wrong chapter(s)? Let us know.